## Summary The filtration performance of breathing system filters can be measured by challenging the filter with sodium chloride particles. However, the challenge itself may alter the filtration performance. This effect was determined by repeating a standard challenge test five times on one sample
The bacterial and viral filtration performance of breathing system filters*
β Scribed by A. R. Wilkes; J. E. Benbough; S. E. Speight; M. Harmer
- Publisher
- John Wiley and Sons
- Year
- 2000
- Tongue
- English
- Weight
- 241 KB
- Volume
- 55
- Category
- Article
- ISSN
- 0003-2409
No coin nor oath required. For personal study only.
β¦ Synopsis
The bacterial and viral filtration performance of 12 breathing system filters was determined using test methods specified in the draft European standard for breathing system filters, BS EN 13328-1. All the filters were of two types, either pleated hydrophobic or electrostatic, and these two types differed in their filtration performance. The filtration performance is expressed in terms of the microbial penetration value, defined as the number of microbes passing through the filter per 10 million microbes in the challenge. The geometric mean (95% confidence limits) microbial penetration value was 1.0 (0.5, 3.5) and 2390 (617, 10 000) for the pleated hydrophobic and electrostatic filters, respectively, for the bacterial challenge, and 87 (48, 212) and 32 600 (10 900, 84 900), respectively, for the viral challenge. In general, there was little change in the microbial penetration values following 24 h simulated use. It is concluded that results from the tests specified in the draft standard will allow comparisons to be made between different manufacturers' products enabling an informed choice to be made.
π SIMILAR VOLUMES
The filtration performance of 33 breathing system filters (nine pleated hydrophobic and 24 electrostatic filters) was measured using sodium chloride particles. The particles had a size distribution with a count median diameter of 0.07βΒ΅m and a geometric standard deviation not exceeding 1.83. The geo
A breathing system filter can remain connected to a laryngeal mask airway during recovery from anaesthesia. The Luerβlock monitoring port on the filter could be used for the delivery of supplementary oxygen. The efficacy of this technique was assessed __in vitro__ by comparing the performance of two
The ability of 29 different breathing system filters (five pleated hydrophobic and 24 electrostatic) to prevent the passage of water was assessed. Five, 10 or 20βml of water was added on to the patient side of the filter during ventilation of a patient model with the filter layer orientated either h