โฆ Synopsis
The characteristics of four commercial strain-gauge pressure transducers suitable for measurements in a temperature range from 4.2 to 293 K were determined in the presence of magnetic fields up to 6 T The highest pressure measured was 3.5 MPa. Tests showed that if the transducers are used in these conditions and no corrections are made, uncertainties up to 19% of the transducer full scale may be inherent in measurements. When such high error levels are unacceptable, the transducers must be individually calibrated under the working conditions: in this way the uncertainties in pressure measurements can usually be kept below 1% of the transducer full scale.
Testing of strain-gauge pressure transducers up to 3.5 MPa at cryogenic temperatures and in magnetic fields up to 6 T
๐ SIMILAR VOLUMES
Comprehensive ( p, ฯ, T ) measurements on pure sulphur hexafluoride have been carried out in the single-phase region (609 values) and along the entire (vapour + liquid) coexistence curve (95 values). The results reported in this first paper cover the homogeneous gas and liquid region in the temperat
Measurements of ( p, ฯ, T ) for {x NH 3 + (1 -x)H 2 O} at x = (1.0000, 0.8374, 0.6005, and 0.2973) and at specified temperatures and pressures in the compressed liquid phase were carried out with a metal-bellows variable volumometer between T = 310 K and T = 400 K at pressures up to 17 MPa. The resu
p, ฯ, T ) measurements of (ammonia + water) in the temperature range 310 K to 400 K at pressures up to 17 MPa. I. Measurements of {xNH 3 + (1 -x)H 2 O} at x = (0.8952, 0.8041, 0.6985, 0.5052, and 0.1016)