In hip replacement the wear rate is reduced when using femoral heads and acetabular cups made of alumina ceramics. It is proposed to use Y-TZP zirconia ceramics. The process on the articulating surfaces head/cup are not totally understood, especially the temperature rise due to frictional heating. Alumina is a stable material, no problems are expected. When using Y-TZP its phase transition may be triggered due to frictional heating and the leaching out of yttria.
The model proposed by Blok, Jaeger and Kuhlmann-Wilsdorf was used to calculate the flash temperature under in vivo conditions. The surface temperature of the wear couple Y-TZP/Y-TZP was higher than the one for alumina/alumina.