𝔖 Bobbio Scriptorium
✦   LIBER   ✦

T2-weighted spine imaging with a fast three-point dixon technique: Comparison with chemical shift selective fat suppression

✍ Scribed by Jingfei Ma; Sanjay K. Singh; Ashok J. Kumar; Norman E. Leeds; Jiping Zhan


Publisher
John Wiley and Sons
Year
2004
Tongue
English
Weight
232 KB
Volume
20
Category
Article
ISSN
1053-1807

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Purpose

To develop a phased‐array coil‐compatible, fast three‐point Dixon (TPD) technique, and compare its performance in T2‐weighted spine imaging with that of the standard chemical shift selective (CHESS) fat suppression technique.

Materials and Methods

We acquired T2‐weighted spine images of 27 patients using essentially identical scanning parameters with the fast TPD technique and standard fast spin echo (FSE) with CHESS fat suppression. A phased‐array coil‐compatible image reconstruction algorithm was developed to generate separate water and fat images from the data acquired with the fast TPD technique. Three neuroradiologists independently scored the images from the two different techniques for uniformity of fat suppression and lesion conspicuity using a four‐point system (1 = poor, 2 = fair, 3 = good, 4 = best).

Results

The reviewers' mean scores were 3.2 and 2.1 for the uniformity of fat suppression, and 3.0 and 2.0 for the lesion conspicuity for the fast TPD and the CHESS fat suppression techniques, respectively. The fast TPD technique was statistically superior to the CHESS technique at P < 0.0005.

Conclusion

The fast TPD technique provides superior fat suppression and lesion conspicuity, and potentially can be used as an alternative to T2‐weighted imaging of the spine. J. Magn. Reson. Imaging 2004;20:1025–1029. © 2004 Wiley‐Liss, Inc.