Susceptibility weighted imaging with multiple echoes
✍ Scribed by Christian Denk; Alexander Rauscher
- Publisher
- John Wiley and Sons
- Year
- 2009
- Tongue
- English
- Weight
- 371 KB
- Volume
- 31
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Purpose:
To extend susceptibility weighted imaging (SWI) to multiple echoes with an adapted homodyne filtering of phase images for the computation of venograms with improved signal to noise ratio (SNR) and contrast to noise ratio (CNR) and to produce high resolution maps of R~2~* relaxation.
Materials and Methods:
Three‐dimensional multi echo gradient echo data were acquired with five equidistant echoes ranging from 13 to 41 ms. The phase images of each echo were filtered with filter parameters adjusted to the echo time, converted into a phase mask, and combined with the corresponding magnitude images to obtain susceptibility weighted images. The individual images were then averaged. Conventional single echo data were acquired for comparison. Maps of R~2~* relaxation rates were computed from the magnitude data. Field maps derived from the phase data were used to correct R~2~* for the influences from background inhomogeneities of the static magnetic field.
Results:
Compared with the single echo images, the combined images had an increase in SNR by 46% and an improvement in CNR by 34 to 80%, improved visibility of small venous vessels and reduced blurring along the readout direction. The R~2~* values of different tissue types are in good agreement with values from the literature.
Conclusion:
Acquisition of SWI with multiple echoes leads to an increase in SNR and CNR and it allows the computation of high resolution maps of R~2~* relaxation. J. Magn. Reson. Imaging 2010;31:185–191. © 2009 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract Susceptibility‐weighted imaging (SWI) consists of using both magnitude and phase images from a high‐resolution, three‐dimensional, fully velocity compensated gradient‐echo sequence. Postprocessing is applied to the magnitude image by means of a phase mask to increase the conspicuity of
## Abstract ## Purpose To investigate whether the variable forms of putative iron deposition seen with susceptibility weighted imaging (SWI) will lead to a set of multiple sclerosis (MS) lesion characteristics different than that seen in conventional MR imaging. ## Materials and Methods Twenty‐s
## Abstract It is shown that a repetitive pulse sequence consisting of two 90° pulses and gradients in a 1:2 ratio around the second 90° pulse generates interscan shifted stimulated echoes (SSTEs) and intrascan multiple spin echoes (MSEs). Separation of these two types of signals is accomplished us
The ornamental design for a magnetic resonance imaging scanner, as shown.
## Abstract A new method is proposed for reducing the artifacts produced in diffusion‐weighted imaging. When data are acquired using multiple shot echo planar acquisitions, conventional reconstruction methods produce artifactual images as a consequence of diffusion weighting and small amounts of bu