Stabilization of ferrocene leakage by physical retention in a cellulose acetate membrane. The fructose biosensor
✍ Scribed by Ján Tkáč; Igor Voštiar; Peter Gemeiner; Ernest Šturdı́k
- Book ID
- 104443879
- Publisher
- Elsevier Science
- Year
- 2002
- Tongue
- English
- Weight
- 96 KB
- Volume
- 55
- Category
- Article
- ISSN
- 1567-5394
No coin nor oath required. For personal study only.
✦ Synopsis
The prevention of ferrocene leakage from an electrode by physical retention of mediator in a matrix of cellulose acetate membrane is reported. Five types of the cellulose acetate membranes were prepared, containing 1.8%, 5.3%, 8.5%, 20.0% of ferrocene and a membrane containing 1.8% of ferrocene and 0.05 % of NafionR in the matrix. Ferrocene embedded membranes were successfully applied in the construction of a fructose biosensor by immobilization of PQQ-dependent fructose dehydrogenase (FDH). The biosensor comprising a cellulose acetate membrane with 1.8% of ferrocene and 0.05% of NafionR had good stability characteristics, retained almost 40% of the initial response after 8 h of continuous use with an initial sensitivity of 226 nA mM À 1 and response time of 75 s.