We have developed a rapid, nondestructive technique using energy-dispersive X-ray fluorescence spectroscopy to determine trace element concentrations in samples of archaeological felsite. The method employs a series of secondary targets that are selected to provide optimal sensitivity to groups of e
Simultaneous analysis of trace concentrations of lead and arsenic by energy-dispersive x-ray fluorescence spectrometry
✍ Scribed by Dariusz Wȩgrzynek; Barbara Hołyńska
- Publisher
- Elsevier Science
- Year
- 1993
- Tongue
- English
- Weight
- 341 KB
- Volume
- 44
- Category
- Article
- ISSN
- 0969-8043
No coin nor oath required. For personal study only.
✦ Synopsis
A method for the analysis of trace concentrations of lead by Energy-Dispersive x-Ray Fluorescence spectrometry when arsenic is also present in the sample is described. To apply the method, the PbLp-line, emitted from the sample to be analyzed, must be available and it cannot interfere with x-ray peaks from other constituents in the sample. An additional reference sample containing lead and no arsenic is required. The intensity ratio (Pb,,/Pb,,) is used for correction using a reference which is arsenic free. By applying the emission-transmission method (Giauque Ed al., 1973) the corrected intensities of Pb,,-and As,,-peaks are calculated for samples of unknown composition.
Based on the corrected intensities the concentrations of Pb and As are determined. To verify the method the IAEA reference material, Marine Sediment SD-M-Z/TM, was analyzed and the results are presented here.
📜 SIMILAR VOLUMES
Trace elemental analyses of rubies and sapphires from different countries of origin were carried out by the EDXRF technique. It was found that K, Ti, Cr, Fe, Cu, Zn, Sr and Ba are generally present in natural rubies and Cr, Ni, Cu and Zr in synthetic ruby. The synthetic ruby shows a clean spectrum d