Simulation of scanning capacitance microscopy measurements on micro-sectioned and bevelled n+–p samples
✍ Scribed by L. Ciampolini; F. Giannazzo; M. Ciappa; W. Fichtner; V. Raineri
- Publisher
- Elsevier Science
- Year
- 2001
- Tongue
- English
- Weight
- 128 KB
- Volume
- 4
- Category
- Article
- ISSN
- 1369-8001
No coin nor oath required. For personal study only.
✦ Synopsis
In this work we compare simulations of SCM measurements on bevelled and micro-sectioned samples for investigating the impact of the carrier spilling effect. Simulation results are validated by experimental data obtained from dedicated samples calibrated by spreading resistance profiling. We show that 3D-based direct inversion of data measured on micro-sectioned bipolar samples results into a good quantitative agreement with the experimental spreading resistance profile in the area where carrier spilling is negligible. 2D-based simulations of bevelled bipolar samples reproduce qualitatively the measured SCM profiles. These results lead to the conclusion that also in the case of SCM, the lateral resolution in junction delineation can be improved by characterizing bevelled samples.