The 4p NaI(Tl) detectors are usually utilized for low-level radioactivity measurements of environmental samples in gamma-ray spectroscopy due to the near 4p solid angle and high full-energy peak and total efficiencies. An analytical theoretical approach is presented here and used to calibrate a 4p N
✦ LIBER ✦
Simple analytical formula to calculate γ-ray cylindrical detectors efficiencies
✍ Scribed by A. Hamzawy
- Publisher
- Elsevier Science
- Year
- 2010
- Tongue
- English
- Weight
- 462 KB
- Volume
- 624
- Category
- Article
- ISSN
- 0168-9002
No coin nor oath required. For personal study only.
✦ Synopsis
In this paper, we introduce a new simplified and reduced expressions for a direct analytical mathematical method to calculate the efficiencies of gamma cylindrical detectors with an arbitrarily positioned point source. The expressions are deduced through a simple and easy straightforward mathematical manner. The expressions are extended to include a co-axial disc source. The current theoretical and the published efficiency values are in good agreement.