Lead selective electrodes based on a di-and tetrathioamide functionalized calix [ 41 arene as ionophores were investigated. The Pb( II)-response functions exhibited almost theoretical Nernstian slopes in the activity range 10-6-10-2 M of lead ions. For both ionophores a preference for lead over othe
Silver selective electrodes based on thioether functionalized calix[4]arenes as ionophores
✍ Scribed by Elżbieta Malinowska; Zbigniew Brzózka; Krzysztof Kasiura; Richard J.M. Egberink; David N. Reinhoudt
- Publisher
- Elsevier Science
- Year
- 1994
- Tongue
- English
- Weight
- 728 KB
- Volume
- 298
- Category
- Article
- ISSN
- 0003-2670
No coin nor oath required. For personal study only.
✦ Synopsis
Silver selective eiectrodes based on thioether functianaiized calix [ 4j arenes 1 and 2 as ionophores were investigated. For both ionophores the selectivity coefficients (log k,,,) were lower than -2.2 for Hg(II) and lower than -4.6 for other cations tested. The best results were obtained with membranes containing dithioether fimctionalized calix[4]arene (ionophore 2), potassium tetrakis( 4-chlorophenyl)borate (KTpClFB) and bis( l-butylpentyl)adipate (BBPA) as a plasticizer. The Ag( I)response functions exhibited almost theoreticd Nernstian slopes in the activity range 1O-6-1O-1 M of silver ions.
📜 SIMILAR VOLUMES
Two calix[4] arene derivatives, in the partial cone conformation, with sulfur-containing functionalities, were tested as neutral carrier ionophores in potentiometric silver-selective electrodes of conventional membrane and membrane-coated glassy carbon electrode types. Comparison with a calix[4]are