## Abstract The solid sample inhomogeneity is a weak point of traditional MALDI deposition techniques that reflects negatively on quantitative analysis. The recently developed sieve‐based device (SBD) sample deposition method, based on the electrospraying of matrix/analyte solutions through a groun
Sieve-based device for MALDI sample preparation. II. Instrumental parameterization
✍ Scribed by Simone Cristoni; Laura Molin; Sara Crotti; Luigi Rossi Bernardi; Roberta Seraglia; Pietro Traldi
- Publisher
- John Wiley and Sons
- Year
- 2009
- Tongue
- English
- Weight
- 349 KB
- Volume
- 44
- Category
- Article
- ISSN
- 1076-5174
- DOI
- 10.1002/jms.1637
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
A simple device effective for matrix‐assisted laser desorption/ionization (MALDI) sample preparation, based on the spraying of matrix/sample solution through a stainless steel sieve (sieve‐based device, SBD), has been employed for the preparation of MALDI samples of peptides, polysaccharides and high molecular weight proteins. The instrumental parameters, mainly the sprayer voltage, solution and auxiliary gas flows, and distances between sprayer–sieve and sieve–MALDI sample holder have been optimized. The MALDI spectra obtained by laser irradiation of the microspots so obtained, exhibit resolution and sensitivity higher than those achieved by the commonly employed dried droplets (DD) method. Worth noting is the fact that, in contrast to what is observed in the DD method, identical spectra are obtained by irradiation of different areas of the MALDI sample, showing that by using the SBD the matrix/analyte ratio is constant over the entire surface. Copyright © 2009 John Wiley & Sons, Ltd.
📜 SIMILAR VOLUMES
## Abstract Spraying of oligonucleotide‐matrix solutions through a stainless steel (ss) sieve (38 µm, 450 mesh) leads to the formation, on the matrix‐assisted laser desorption/ionization (MALDI) sample holder, of uniformly distributed microcrystals, well separated from each other. When the resultin