Short data-acquisition times improve projection images of lung tissue
✍ Scribed by Dean O. Kuethe; Natalie L. Adolphi; Eiichi Fukushima
- Book ID
- 102953955
- Publisher
- John Wiley and Sons
- Year
- 2007
- Tongue
- English
- Weight
- 891 KB
- Volume
- 57
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
MR images of laboratory rat lungs that resolve the thin membranes that separate lung lobes are presented. It appears that the capabilities of in vivo small‐animal pulmonary MRI may rival those of in vivo small‐animal X‐ray CT. Free induction decay (FID)‐projection imaging was employed with particular attention to the choice of acquisition time. For a given nominal resolution, one obtains optimal point discrimination when the acquisition time T~acq~ normalized by the signal decay time constant T is approximately 0.8–0.9, although a better signal‐to‐noise ratio (SNR) is obtained when this quotient is 1.6. Currently available equipment should be able to even exceed the results presented herein. Magn Reson Med 57:1058–1064, 2007. © 2007 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
ultiphoton laser scanning microscopy offers improved axial resolution over confocal microscopy for three-dimensional (3D) imaging of fluorescently labeled cells. It also greatly reduces phototoxicity and photobleaching outside the plane of focus when collecting 3D series of optical sections in livin
A fast, reproducible, and noninvasive method is required for quantifying cartilage thickness clinically and for studying the deformation of articular cartilage during and after mechanical loading in vivo. The objective of the current investigation was to test the repeatability of regional distributi