𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Separation of some neutral lipids by normal-phase high-performance liquid chromatography on a cyanopropyl column: Ubiquinone, dolichol, and cholesterol levels in sheep liver

✍ Scribed by David N. Palmer; Maureen A. Anderson; Robert D. Jolly


Publisher
Elsevier Science
Year
1984
Tongue
English
Weight
381 KB
Volume
140
Category
Article
ISSN
0003-2697

No coin nor oath required. For personal study only.

✦ Synopsis


The normal-phase high-performance liquid chromatographic separation of neutral lipids into molecular classes was carried out on a cyanopropyl (CN) column eluted with isopropanol in hexane. Cholesteryl, retinyl, and dolichyl esters, triglycerides and vitamin E, ubiquinone, dolichol, phytol, and cholesterol eluted as separate peaks with 0.05% isopropanol in hexane. Cholesterol, retinol, diglyceride, and monoglyceride eluted as separate peaks with 0.75% isopropanol in hexane. These separations could not be achieved on a silica gel column. The method was used to assay sheep liver ubiquinone, dolichol, and cholesterol levels, that were determined as 77, 108 and 1864 micrograms/g wet wt, respectively.