## Abstract A self‐calibrated parallel imaging reconstruction method is proposed for azimuthally undersampled radial dataset. A generalized auto‐calibrating partially parallel acquisition (GRAPPA) operator is used to widen each radial view into a band consisting of several parallel lines, followed
Self-calibrating GRAPPA operator gridding for radial and spiral trajectories
✍ Scribed by Nicole Seiberlich; Felix Breuer; Martin Blaimer; Peter Jakob; Mark Griswold
- Publisher
- John Wiley and Sons
- Year
- 2008
- Tongue
- English
- Weight
- 290 KB
- Volume
- 59
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Self‐calibrating GRAPPA operator gridding (GROG) is a method by which non‐Cartesian MRI data can be gridded using spatial information from a multichannel coil array without the need for an additional calibration dataset. Using self‐calibrating GROG, the non‐Cartesian datapoints are shifted to nearby k‐space locations using parallel imaging weight sets determined from the datapoints themselves. GROG employs the GRAPPA Operator, a special formulation of the general reconstruction method GRAPPA, to perform these shifts. Although GROG can be used to grid undersampled datasets, it is important to note that this method uses parallel imaging only for gridding, and not to reconstruct artifact‐free images from undersampled data. The innovation introduced here, namely, self‐calibrating GROG, allows the shift operators to be calculated directly out of the non‐Cartesian data themselves. This eliminates the need for an additional calibration dataset, which reduces the imaging time and also makes the GROG reconstruction more robust by removing possible inconsistencies between the calibration and non‐Cartesian datasets. Simulated and in vivo examples of radial and spiral datasets gridded using self‐calibrating GROG are compared to images gridded using the standard method of convolution gridding. Magn Reson Med 59:930–935, 2008. © 2008 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract A fast parallel MRI (pMRI) reconstruction method is presented for 2D and 3D radial trajectories. A limitation of the radial generalized autocalibrating partially parallel acquisitions (GRAPPA) method is the need to acquire training data prior to the actual scan. This can be eliminated b