## Abstract Patellofemoral dysfunction following total knee replacement (TKR) is a significant clinical problem, but little information exists on the mechanics of the patellofemoral retinacula or the effects of TKR on these structures. We hypothesized that TKR would cause significant elongation of
Retrieval, experimental, and computational assessment of the performance of total knee replacements
β Scribed by Jeremy J. Rawlinson; Bridgette D. Furman; Stephen Li; Timothy M. Wright; Donald L. Bartel
- Publisher
- Elsevier Science
- Year
- 2006
- Tongue
- English
- Weight
- 443 KB
- Volume
- 24
- Category
- Article
- ISSN
- 0736-0266
No coin nor oath required. For personal study only.
β¦ Synopsis
Wear mechanisms in polyethylene components for total knee replacements are inherently mechanical; the local stresses or strains exceed some material limit. Retrieval analysis and knee simulators have provided the means to quantify the damage observed in vivo or in vitro. These results have been circumstantially linked to the material stresses obtained from computational simulations using finite element analysis, knee simulator tests, and computational simulations of two condylar knee designs. We hypothesize that if an equivalent loading environment is produced in the computational simulation, we can correlate the distribution of computed stresses with observed damage of simulator specimens and further relate design differences to in vivo performance from retrieval analyses. The finite element model agreed with the knee simulator kinematics and kinetics within 2-13%, and composite FEA contact areas matched 66-90% of the damage areas due to burnishing on the simulator specimens. Burnishing was the primary mode of damage for both the simulator and retrieval specimens corresponding with the relatively low magnitudes of contact stress observed. Both the computational and experimental techniques underpredicted the amount determined from retrieval analysis, but the differences between the two designs were consistent for all three methods. Combining these techniques strengthens the applicability of the computational simulation while highlighting the complementary approach of these methods for preclinical testing and assessing the link between material state and damage.
π SIMILAR VOLUMES
## Abstract Thirtyβeight cemented acetabular components that had been clinically implanted in clientβowned dogs were retrieved postmortem and analyzed for mechanical stability, volumetric wear, and articular surface damage. Comparison of the results from this study with similar studies on autopsyβr
## Abstract ## Objective To compare the EuroQol (EQβ5D) and Short Form 6D (SFβ6D) among multiethnic Asian patients with knee osteoarthritis (OA) scheduled for total knee replacement in Singapore. ## Methods Patients were asked to complete questionnaires including the EQβ5D, Short Form 36, Wester
Due to their desirable broadband noise attenuation characteristics, expansion chambers are widely used in the ducting systems for pulsating flows, including the breathing systems of engines and reciprocating turbomachinery. The present study investigates in detail the effect of the length on the aco