𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Reserve nitrous oxide cylinders on anaesthetic machines : apparatus : A survey of attitudes and equipment at a large DGH

✍ Scribed by G. F. Rousseau; A. S. Carr


Publisher
John Wiley and Sons
Year
2000
Tongue
English
Weight
197 KB
Volume
55
Category
Article
ISSN
0003-2409

No coin nor oath required. For personal study only.

✦ Synopsis


The presence of reserve nitrous oxide cylinders on anaesthetic machines is not necessary and potentially hazardous, wasteful and polluting. Although reserve nitrous oxide cylinders are routinely present on the anaesthetic machines in many hospitals in the UK, it is our concern that they are not checked or maintained properly. We examined a sample of cylinders and surveyed the practice amongst the anaesthetic staff in our Department with an anonymous questionnaire. This confirmed that: (1) the equipment is not properly maintained, 68% of cylinders sampled having passed a 3‐year expiry date; (2) checks are of little practical use, are not routinely performed by anaesthetists, and only 46% of respondents routinely check the nitrous oxide cylinders. In conclusion, the nitrous oxide cylinders are not properly checked or maintained and may represent a safety hazard.