Reproducibility of 3.0 Tesla magnetic resonance spectroscopy for measuring hepatic fat content
✍ Scribed by Jochem R. van Werven; Johannes M. Hoogduin; Aart J. Nederveen; Andre A. van Vliet; Ewa Wajs; Petra Vandenberk; Erik S.G. Stroes; Jaap Stoker
- Publisher
- John Wiley and Sons
- Year
- 2009
- Tongue
- English
- Weight
- 324 KB
- Volume
- 30
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Purpose
To investigate reproducibility of proton magnetic resonance spectroscopy (^1^H‐MRS) to measure hepatic triglyceride content (HTGC).
Materials and Methods
In 24 subjects, HTGC was evaluated using ^1^H‐MRS at 3.0 Tesla. We studied “between‐weeks” reproducibility and reproducibility of ^1^H‐MRS in subjects with fatty liver. We also studied within liver variability and within day reproducibility. Reproducibility was assessed by coefficient of variation (CV), repeatability coefficient (RC), and intraclass correlation coefficient (ICC).
Results
The CV of between weeks reproducibility was 9.5%, with a RC of 1.3% HTGC (ICC 0.998). The CV in fatty livers was 4.1%, with a RC of 1.3% HTGC (ICC 0.997). Within day CV was 4.5%, with a RC of 0.4% HTGC (ICC 0.999). CV for within liver variability was 14.5%.
Conclusion
Reproducibility of ^1^H‐MRS to measure HTGC for “between‐weeks” measurements and in fatty livers is high, which is important for follow‐up studies. Within liver variability displays a larger variation, meaning that liver fat is not equally distributed and during consecutive measurements the same voxel position should be used. J. Magn. Reson. Imaging 2009;30:444–448. © 2009 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract ## Purpose: To determine the reproducibility of proton MR spectroscopy (^1^H‐MRS) for assessing vertebral bone marrow adiposity at 3 Tesla (T); to evaluate variation of marrow adiposity at different vertebral levels; and to demonstrate the feasibility of using ^1^H‐MRS at 3T for evalua
A new theoretical method is presented for designing frequency responses of double-tuned, low-pass birdcage coils. This method is based on Kirchhoff's equations through a nonsymmetric matrix algorithm and extended through a modification of the corresponding eigenvalue system from a single-tuned mode.
## Abstract ## Purpose: To assess the accuracy of noninvasive 3.0 T ^1^H‐magnetic resonance spectroscopy (^1^H‐MRS) in an experimental steatosis model for the discrimination of clinically relevant macrovesicular steatosis degrees and to evaluate three different ^1^H‐MR spectrum‐based fat quantific
## Abstract ## Purpose: To compare correlations of intramyocellular lipids (IMCL) measured by short and long echo‐time proton magnetic resonance spectroscopy (1H‐MRS) with indices of body composition and insulin resistance in obese and normal‐weight women. ## Materials and Methods: We quantified
## Abstract ## Purpose To investigate the value of in vivo proton magnetic resonance spectroscopy (MRS) in the assessment of large focal hepatic lesions and to measure the metabolite change of hepatocellular carcinoma (HCC) after transcatheter arterial chemoembolization (TACE) using 3.0‐T scanner.