Reproducibility and dependence on diffusion weighting of line scan diffusion in the lumbar intervertebral discs

✍ Scribed by David C. Newitt; Sharmila Majumdar


Publisher
John Wiley and Sons
Year
2005
Tongue
English
Weight
507 KB
Volume
21
Category
Article
ISSN
1053-1807

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Purpose

To investigate the dependence of line scan diffusion imaging (LSDI) in the lumbar vertebral discs on diffusion weighting, fat suppression (FS), and postprocessing noise correction.

Materials and Methods

Eleven normal volunteers were scanned using 4 b‐value and 12 b‐value LSDI protocols, with and without FS. Three repeated four b‐value scans were performed for evaluation of the reproducibility of apparent diffusion coefficient (ADC) values calculated with mono‐ and biexponential decay models. Two‐point ADC analysis for 12 b‐value scans was performed with and without noise correction to evaluate the ADC dependence on diffusion weighting. Correlations between different ADC calculation and acquisition methods were evaluated.

Results

Monoexponential ADC measures had a coefficient of variation (CV) under 3%, while use of a constrained biexponential increased the CV to 6% to 9%. Strong dependence on b‐value was seen from chemically shifted marrow fat signal and noise. These systematic variations in ADC were eliminated using noise correction and FS. ADC values from 4 and 12 b‐value FS scans correlated strongly (R^2^ = 0.91), while biexponentially derived ADC values correlated moderately well with the FS ADC (R^2^ = 0.51).

Conclusion

LSDI gives reproducible ADC measurements in the lumbar discs, largely independent of b‐value and signal‐to‐noise ratio (SNR) when used with noise correction and FS. J. Magn. Reson. Imaging 2005;21:482–488. © 2005 Wiley‐Liss, Inc.