๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Reliability and reproducibility of automated contour analysis in intravascular ultrasound images of femoropopliteal arteries

โœ Scribed by Aad van der Lugt; Aran Hartlooper; Jerden A. van Essen; Wenguang Li; Clemens von Birgelen; Johan H.C. Reiber; Elma J. Gussenhoven


Publisher
Elsevier Science
Year
1998
Tongue
English
Weight
755 KB
Volume
24
Category
Article
ISSN
0301-5629

No coin nor oath required. For personal study only.

โœฆ Synopsis


An automated contour analysis system was previously developed to increase reproducibility and facilitate quantitative analyses of intravascular ultrasound (IVUS) images. The aim of this study was to compare measurements by this automated system with those obtained by conventional manual tracing, and to determine the intra- and interobserver variability of the automated system. IVUS images obtained in the femoropopliteal artery (n = 12) were analyzed with both systems. Area measurements by the automated system agreed well with the results obtained by manual tracing, displaying low coefficients of variation (8.5 to 15.7%) and high correlation coefficients (r = 0.92 to 0.98). Intra- and interobserver comparison of lumen area, vessel area, plaque area and percentage area stenosis showed low coefficients of variation (6.0 to 15.3% and 5.7 to 14.0%, respectively) and high correlation coefficients (both: r = 0.93 to 0.99). These data indicate that the automated analysis system is a reliable tool for the quantitative assessment of vessel dimensions in IVUS images obtained during clinical examination of peripheral arteries.


๐Ÿ“œ SIMILAR VOLUMES


Analysis of reproducibility of reference
โœ Kearney, Peter P. ;Ramo, M. Pauliina ;Shaw, T. R. D. ;Starkey, Ian R. ;McMurray, ๐Ÿ“‚ Article ๐Ÿ“… 1997 ๐Ÿ› John Wiley and Sons ๐ŸŒ English โš– 40 KB ๐Ÿ‘ 2 views

Intravascular ultrasound is widely used to guide coronary stent implantation. The key quantitative criterion for successful implantation is the demonstration of adequate expansion of the stented lumen relative to that of the adjacent reference vessel segments. In this study we aimed to establish the