## Abstract We report noninvasive imaging of the brain of adult Zebrafish (__Danio rerio__) using real time optical coherence tomography (OCT) capable of acquiring cross sectional 2D OCT images @ 8 frames/sec. Anatomic features such as telencephalon, tectum opticum, eminentia Granularis and cerebel
Real-time tracking of vocal fold injections with optical coherence tomography
β Scribed by James A. Burns; Ki Hean Kim; James B. Kobler; Johannes F. deBoer; Gerardo Lopez-Guerra; Steven M. Zeitels
- Publisher
- John Wiley and Sons
- Year
- 2009
- Tongue
- English
- Weight
- 450 KB
- Volume
- 119
- Category
- Article
- ISSN
- 0023-852X
No coin nor oath required. For personal study only.
β¦ Synopsis
Abstract
Objectives/Hypothesis:
Optical coherence tomography (OCT) is a promising new imaging modality that can help discern the layered microstructure of vocal folds. In the future, subepithelial injections of implants will improve vocal fold pliability where there is stiffness of phonatory mucosa. Using OCT both to delineate the depth of subepithelial injections realβtime and to serially image the implant over time would be valuable, and has not been demonstrated previously.
Study Design:
Ex vivo study using excised calf larynges and survival study using canines in vivo.
Methods:
An investigation was done employing realβtime OCT imaging of subepithelial injection pulses into phonatory mucosa of four calf larynges ex vivo to track the presence of subepithelial implants in phonatory mucosa in a survival study using an in vivo canine model.
Results:
OCT readily identified polyethylene glycol (PEG)βbased hydrogel in the subepithelial plane in both the ex vivo calf study and the in vivo canine study. Ex vivo calf images correlated with histological specimens obtained immediately postinjection. Images obtained in this study provide confirmation of the hydrogel injection depth in real time, and allow for the implant to be tracked during a canine survival study.
Conclusions:
OCT can confirm subepithelial placement of hydrogel implant in the vocal fold with sufficient resolution to provide instantaneous feedback of an injection pulse. Survival studies in an in vivo canine model indicate that OCT can potentially be helpful in monitoring rheologically appropriate implants within the superficial lamina propria. Laryngoscope, 2009