Real-time geometric distortion correction for interventional imaging with echo-planar imaging (EPI)
✍ Scribed by Iulius Dragonu; Baudouin Denis de Senneville; Bruno Quesson; Chrit Moonen; Mario Ries
- Publisher
- John Wiley and Sons
- Year
- 2009
- Tongue
- English
- Weight
- 349 KB
- Volume
- 61
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Many MR‐guided interventional procedures rely on fast imaging sequences for providing images in real‐time with a precise relation between the target position in the image and its true position. Echo‐planar imaging (EPI) methods are very fast but prone to geometric distortions. Here, we propose a correction method designed for real‐time conditions, adapting existing approaches based on dual EPI acquisition with varying echo times. The method is demonstrated with MR‐thermometry for guiding thermal therapies. The proposed approach imposes a small penalty in acquisition speed but adds negligible latency to data processing, an important element for interventions of mobile organs. Magn Reson Med, 2009. © 2009 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
A novel and effective technique is described for distortion correction in echo planar imaging (EPI) utilizing the field maps derived from multi-echo gradient-echo images. The distortions from different off-resonance related factors such as field inhomogeneity, eddy current effect, radiofrequency pul
## Abstract A new method is described that can correct the distortions due to multiple off‐resonance effects in echo planar imaging, including those caused by __B__~0~ field inhomogeneities, eddy currents, and gradient waveform imperfections. The proposed method uses a phase encoded acquisition and
## Abstract We present, here, a simple method for measurement and correction of off‐resonance related geometric distortion in echo‐planar imaging (EP1). This method uses high signal‐to‐noise ratio (SNR) EPI‐based field maps, rapidly acquired using a series of gradient recalled images collected acro
## Abstract Diffusion‐weighted images acquired with the echo‐planar imaging technique are highly sensitive to eddy current induced geometric distortions that vary with the magnitude and direction of the diffusion sensitizing gradients. Such distortions cause misalignment of images acquired with dif