Rapid quantification of HIV protease inhibitors in human plasma by high-performance liquid chromatography coupled with electrospray ionization tandem mass spectrometry
✍ Scribed by K. M. L. Crommentuyn; H. Rosing; L. G. A. H. Nan-Offeringa; M. J. X. Hillebrand; A. D. R. Huitema; J. H. Beijnen
- Publisher
- John Wiley and Sons
- Year
- 2003
- Tongue
- English
- Weight
- 165 KB
- Volume
- 38
- Category
- Article
- ISSN
- 1076-5174
- DOI
- 10.1002/jms.425
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
HIV protease inhibitors are important antiretroviral drugs which have substantially reduced the morbidity and mortality associated with HIV‐1 infection. Recent data have shown relationships between plasma concentrations of the protease inhibitors and clinical response, which makes therapeutic drug monitoring valuable. We have developed and validated an assay, using liquid chromatography coupled with electrospray tandem mass spectrometry (LC/MS/MS), for the routine quantification of the six licensed protease inhibitors (amprenavir, indinavir, lopinavir, nelfinavir, ritonavir and saquinavir) and the pharmacologically active nelfinavir metabolite M8 in plasma.
The sample pretreatment consisted of protein precipitation with a mixture of methanol and acetronitrile using only 100 µl of plasma. Chromatographic separation was performed on an Inertsil ODS3 column (50 × 2.0 mm i.d., particle size 5 µm), with a quick stepwise gradient using an acetate buffer (pH 5) and methanol, at a flow rate of 0.5 ml min^−1^. The analytical run time was 5.5 min. The use of a 96‐well plate autosampler allowed batch sizes up to 150 patient samples. The triple‐quadrupole mass spectrometer was operated in the positive ion mode and multiple reaction monitoring was used for drug quantification. The method was validated over the concentration ranges 0.01–10 µg ml^−1^ for indinavir and saquinavir, 0.1–10 µg ml^−1^ for amprenavir, 0.05–10 µg ml^−1^ for nelfinavir and ritonavir, 0.1–20 µg ml^−1^ for lopinavir and 0.01–5 µg ml^−1^ for M8. Saquinavir‐d~5~ and indinavir‐d~6~ were used as internal standards. The coefficients of variation were always <10% for both intra‐day and inter‐day precisions for each compound. Mean accuracies were also between the designated limits (±15%). The validated concentration ranges proved to be adequate in daily practice. This robust and fast LC/MS/MS assay is now successfully applied for routine therapeutic drug monitoring and pharmacokinetic studies in our hospital. Copyright © 2003 John Wiley & Sons, Ltd.
📜 SIMILAR VOLUMES
## Abstract A rapid, sensitive and specific method for quantifying clonazepam in human plasma using diazepam as the internal standard (IS) is described. The analyte and the IS were extracted from plasma by liquid‐liquid extraction using a hexane/diethylether (20 : 80, v/v) solution. The extracts we
## Abstract A simple, sensitive and rapid high‐performance liquid chromatography/electrospray ionization tandem mass spectrometry method was developed and validated for the assay of granisetron in human plasma. Following liquid–liquid extraction, the analytes were separated using an isocratic mobil
A method was developed for the bio-analysis of Ecteinascidin 743 (ET-743) using miniaturized liquid chromatography (LC) coupled to an electrospray ionization sample inlet (TurbolonSpray) and two quadrupole mass analyzers (LC/ESI-MS/MS). Solid-phase extraction was used as a sample pretreatment proced
## Abstract A rapid, sensitive and specific analytical method was developed and validated to quantify gabapentin in human plasma using acetaminophen as an internal standard. The method employs a single plasma protein precipitation. The analytes are chromatographed on a C~4~ reversed‐phase chromatog