## Abstract The effects of radiation sterilization on ultrahigh molecular weight polyethylene were explored by freeze fractures, electron spin resonance, absorption of aqueous media, density measurements, smallβangle and wideβangle xβray scatter, differential scanning calorimetry, infrared and ultr
Radiation sterilization and the wear rate of polyethylene
β Scribed by Dr. R. M. Rose; E. V. Goldfarb; E. Ellis; A. N. Crugnola
- Publisher
- Elsevier Science
- Year
- 1984
- Tongue
- English
- Weight
- 729 KB
- Volume
- 2
- Category
- Article
- ISSN
- 0736-0266
No coin nor oath required. For personal study only.
β¦ Synopsis
The resistance to sliding wear was measured in bovine serum for two polyethylenes with molecular weights of about LO5 and lo6 that had been irradiated in air and nitrogen with y-ray dosages up to 20 Mrad. Molecular weight measurements were performed after irradiation as well. Wear generally increased with dosage and contact stress, becoming measurable in many cases only after a critical dose (or stress) was exceeded. The most significant effect noted was that the irradiation changed the pressure dependence of the wear rate. Thus, whether or not a sterilizing (or resterilizing) dose will measurably increase the wear depends on the contact stress and, therefore, on the specific application. The increase in wear rate appeared to be due to a combination of scission and oxidation, suggesting the practical advisability of radiation sterilization under an inert atmosphere, as confirmed by comparative measurements at the higher dosages.
π SIMILAR VOLUMES
Sterilization by ionizing radiation has become, next to ethylene oxide treatment, the most important "cold" sterilization process for medical devices made from plastics. The effects of ionizing radiation on the most important polymer for medical devices, ultra-high molecular-weight polyethylene, are
The adverse biological response to ultra-high molecular weight polyethylene (UHMWPE) wear debris in total joint replacement warrants the need for the material's improved wear resistance. The reaction of oxygen with radiation-induced free radicals within UHMWPE leads to oxidative degradation of the m