Quantitative phase-contrast flow MRI measurements in the presence of a second vessel closely positioned to the examined vessel

✍ Scribed by Kerstin M. Lagerstrand; Barbro Vikhoff-Baaz; Göran Starck; Eva Forssell-Aronsson


Publisher
John Wiley and Sons
Year
2006
Tongue
English
Weight
464 KB
Volume
23
Category
Article
ISSN
1053-1807

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Purpose

To examine the influence of the truncated sampling of k‐space data on the accuracy of phase‐contrast (PC) flow quantifications in the presence of nearby vessels.

Materials and Methods

Computer simulations were performed along with some experimental validations on a flow phantom and a normal subject.

Results

The accuracy of the PC flow quantification decreased when a second vessel was positioned closely to the examined vessel. The effect strongly depended on the peak flow velocity in the second vessel relative to the velocity encoding (venc) level of the MRI acquisition, and on the position and area of the second vessel.

Conclusion

Due to the truncated sampling of the k‐space data, signal leaked from nearby vessels and distorted the PC flow quantification in the examined vessel. The presented results suggest specific experimental conditions to minimize this flow quantification error. J. Magn. Reson. Imaging 2006. © 2006 Wiley‐Liss, Inc.