Quantitative, low cycle, crack initiation fatigue testing of fine wires and CENELEC standard pacing coil

✍ Scribed by Hildebrand, Laurie B. ;Schmidt, John A. ;Prentice, John K. ;Stotts, Larry J.


Publisher
John Wiley and Sons
Year
1999
Tongue
English
Weight
134 KB
Volume
48
Category
Article
ISSN
0021-9304

No coin nor oath required. For personal study only.

✦ Synopsis


Uniaxial fatigue testing was performed on different diameters of fine wires made from MP35N. The fatigue limits of the wires differed from each other based on the diameter of the wire. Multiaxial (shear) fatigue testing was also performed on a benchmark coil used to evaluate the fatigue life of all modern pacemaker leads (the CENELEC standard coil). A computer algorithm was used to quantify the maximum shear stress and strain on the coil. The bend radius, coil diameter, wire diameter, and pitch of the coil all affect the shear stress and strain and therefore the fatigue properties of conductor coils. Based on the analysis presented, it was determined that the portion of the CENELEC standard dealing with fatigue, when used in its present format, is not a valid fatigue test for pacemaker leads.