𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Quality assurance considerations for use of the FluorImager SI∖R and FragmeNT Analysis software

✍ Scribed by Suzanne A. Christ; Richard N. Silbiger; Manju Garg; Susan E. Franson; Gregory P. Toth


Publisher
John Wiley and Sons
Year
2000
Tongue
English
Weight
267 KB
Volume
21
Category
Article
ISSN
0173-0835

No coin nor oath required. For personal study only.

✦ Synopsis


The Fluorlmager SI (FSI) from Molecular Dynamics is one of several scanning instruments available for the detection of fluorescent emissions associated with DNA samples in a variety of matrices (agarose and polyacrylamide gels, membranes and microplates). In our laboratory, we measured the electrophoretic mobility of randomly amplified polymorphic DNA (RAPD) fragments stained with ethidium bromide in agarose using the FSI to scan gels and the associated Molecular Dynamics software (ImageQuaNT, and FragmeNT Analysis) for analysis. Initial scans and analyses resulted in inconsistent band detection across the same gel and across several scans of the same gel. To determine the best types of calibration for the instrument, several factors were considered and then evaluated. Tests of calibration acceptability were also evaluated. Band detection by FragmeNT Analysis was improved following optimization of matrices and parameters used in calibration and experimental scans. In addition, use of software templates for analysis and modifications in the staining procedure, which have resulted in decreased instrument associated variance, are discussed.