𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Protection of fiber function by para-Axial fluid flow in interstitial laser therapy of malignant tumors

✍ Scribed by Kambiz Dowlatshahi; Julee D. Bangert; Michael F. Haklin; Charles K. Rhodes; Ronald S. Weinstein; Steven G. Economou


Publisher
John Wiley and Sons
Year
1990
Tongue
English
Weight
592 KB
Volume
10
Category
Article
ISSN
0196-8092

No coin nor oath required. For personal study only.

✦ Synopsis


In the past, interstitial laser therapy frequently has failed because of the damage to the bare fiber tip due to intense heat generated at the point of contact. Using a rat mammary tumor model, we describe a method of placing a 600 micron fiber inside a gauge 19 needle cannula after its insertion into the tumor. With this device continuous wave NdYAG laser is delivered to the target tumor while 0.9% saline flows para-axially into the tumor. Significant coagulation necrosis was induced with 500 joules at 5 watts, 100 seconds and 1 cc per minute of saline while the needlefiber is pulled out of the tumor by 10 mm. The mean transmission loss after 500 joules was 2% in ten experiments. The tumor edema due to 1.5 ml of saline was transient. We conclude that successful hyperthermic coagulation necrosis by NdYAG laser can be achieved with minimal transmission loss by employing the above technique.