## Abstract ## Purpose To compare high‐resolution T2‐weigthed images of the liver with and without integrated parallel acquisition techniques (iPAT) using either breath‐hold sequences in combination with prospective acquisition motion correction (PACE) or respiratory triggering. ## Materials and
Prospective correction of affine motion for arbitrary MR sequences on a clinical scanner
✍ Scribed by Kay Nehrke; Peter Börnert
- Book ID
- 102533723
- Publisher
- John Wiley and Sons
- Year
- 2005
- Tongue
- English
- Weight
- 290 KB
- Volume
- 54
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
The concept of prospective 3D affine motion correction was generalized, based on the Bloch equations, for signal excitation and sampling using arbitrary MR sequences. The technique was implemented on a clinical MRI scanner for Cartesian, radial, and spiral imaging sequences, as well as for 2D spatially selective RF excitation pulses. A patient‐specific motion model steered by real‐time navigators was employed to account for the additional degrees of freedom provided by the affine motion model. Different navigator concepts (multiple spatial and temporal navigators, quadratic navigators and other motion sensors) were investigated, with the aim of improving the correlation between navigator information and the motion model. Experiments on moving phantoms are presented to prove the technical feasibility of the approach. In vivo experiments on coronary MRA and renal MRI show the potential of the method for cardiac and abdominal applications hampered by respiratory motion. Magn Reson Med, 2005. © 2005 Wiley‐Liss, Inc.