PROPELLER EPI: An MRI technique suitable for diffusion tensor imaging at high field strength with reduced geometric distortions

✍ Scribed by Fu-Nien Wang; Teng-Yi Huang; Fa-Hsuan Lin; Tzu-Chao Chuang; Nan-Kuei Chen; Hsiao-Wen Chung; Cheng-Yu Chen; Kenneth K. Kwong


Publisher
John Wiley and Sons
Year
2005
Tongue
English
Weight
675 KB
Volume
54
Category
Article
ISSN
0740-3194

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

A technique suitable for diffusion tensor imaging (DTI) at high field strengths is presented in this work. The method is based on a periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) k‐space trajectory using EPI as the signal readout module, and hence is dubbed PROPELLER EPI. The implementation of PROPELLER EPI included a series of correction schemes to reduce possible errors associated with the intrinsically higher sensitivity of EPI to off‐resonance effects. Experimental results on a 3.0 Tesla MR system showed that the PROPELLER EPI images exhibit substantially reduced geometric distortions compared with single‐shot EPI, at a much lower RF specific absorption rate (SAR) than the original version of the PROPELLER fast spin‐echo (FSE) technique. For DTI, the self‐navigated phase‐correction capability of the PROPELLER EPI sequence was shown to be effective for in vivo imaging. A higher signal‐to‐noise ratio (SNR) compared to single‐shot EPI at an identical total scan time was achieved, which is advantageous for routine DTI applications in clinical practice. Magn Reson Med, 2005. © 2005 Wiley‐Liss, Inc.