Probe with chest shielding for improved breast MRI
โ Scribed by Boguslaw Tomanek; David I. Hoult; Xiaoman Chen; Richard Gordon
- Publisher
- John Wiley and Sons
- Year
- 2000
- Tongue
- English
- Weight
- 167 KB
- Volume
- 43
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
โฆ Synopsis
The design and construction of an RF coil system for use in MR breast imaging is described. The two-ring, tuned Helmholtz coil, with its axis perpendicular to the chest, surrounds a single pendant breast and is coupled both internally and to the MRI transmitter/receiver by mutual induction. The addition of two symmetrical RF shields minimizes losses in the chest and significantly improves performance. Images obtained from eight healthy volunteers showed that the coil permitted imaging of breasts of diverse size with an in-plane resolution of 0.27 ุ 0.27 mm and a slice thickness of less than 2 mm at a field strength of 3 T as well as 1.5 T. The use of shields with surface coils in general is advocated as a method for improving signal-to-noise ratio.
๐ SIMILAR VOLUMES
Background and Objectives: Chest wall recurrence occurs in 5-20% of breast cancer patients. Until recently, the only treatments available were surgical resection or radiotherapy. Photodynamic therapy (PDT) is a new modality that uses a photosensitizer and light to destroy tumor cells selectively. We