## Abstract The effects of the pressure drop across the column on retention and efficiency in SFC have been studied. Numerical methods are described which enable the prediction of holdβup time and pressure drop in both packed and open tubular columns. Predictions of both holdβup time and pressure d
Pressure drop effects on selectivity and resolution in packed column supercritical fluid chromatography
β Scribed by Lou, Xianwen ;Janssen, Hans-Gerd ;Snijders, Henri ;Cramers, Carel A.
- Publisher
- John Wiley and Sons
- Year
- 1996
- Tongue
- English
- Weight
- 718 KB
- Volume
- 19
- Category
- Article
- ISSN
- 0935-6304
No coin nor oath required. For personal study only.
β¦ Synopsis
Abstract
The influence of pressure drop on retention, selectivity, plate height and resolution was investigated systematically in packed supercritical fluid chromatography (SFC) using pure carbon dioxide as the mobile phase. Numerical methods developed previously which enabled the prediction of pressure gradients, diffusivities, capacity factors, plate heights and resolutions along the length of the column were used for the model calculations. The effects of inlet pressure and supercritical fluid flow rate on selectivity and resolution are studied. In packed column SFC with pure carbon dioxide as the mobile phase, the pressure drop can have a significant effect on resolution. The flow rate is shown to have a larger effect than generally realized. The calculated data are shown to be in good agreement with the experimental results. Finally, the variation of the chromatographic parameters along a 5.5 meter long model SFC column is illustrated. The possibilities and limitations of using long packed columns in SFC are discussed. It is demonstrated that long columns with large plate numbers do not necessarily yield better separations.
π SIMILAR VOLUMES
## Abstract The retention behavior of polycyclic aromatic hydrocarbons (PAHs) in packedβcolumn supercritical fluid chromatography (SFC) is studied for monomeric and polymeric C~18~ columns. Molecular shape discrimination (shape selectivity) is assessed through the use of Standard Reference Material
## Abstract It is often observed that peak area precision in packed column supercritical fluid chromatography (pSFC) is rather poor when compared to values obtained in HPLC. Since the valves used for sample introduction in both techniques are identical, it is reasonable to expect that injection pre
## High resolution packed column supercritical fluid chromatography of fatty acid methyl esters A wide range of methyl esters of saturated and unsaturated fatty acids from C 16 to C 24 has been separated by supercritical fluid chromatography on a packed silica column. The retention times were prim
The application of fluid extraction in combination with fluid chromatography with packed column and flame ionization detection is described. Fluid chromatographic equipment is shown. Applications of this system to drug characterization are demonstrated.
Fused silica capillary columns with internal diameters from 100 to 25 pm werecoated withSE-54 and evaluated undersupercritical fluid chromatographic conditions using carbon dioxide as mobile phase.Experimental results compared well with theoretical predictions. At ten times the optimum mobile phase