๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Preparation of neutron-activated xenon for liquid xenon detector calibration

โœ Scribed by K. Ni; R. Hasty; T.M. Wongjirad; L. Kastens; A. Manzur; D.N. McKinsey


Publisher
Elsevier Science
Year
2007
Tongue
English
Weight
675 KB
Volume
582
Category
Article
ISSN
0168-9002

No coin nor oath required. For personal study only.

โœฆ Synopsis


We report the preparation of neutron-activated xenon for the calibration of liquid xenon (LXe) detectors. Gamma rays from the decay of xenon metastable states, produced by fast neutron activation, were detected and their activities measured in a LXe scintillation detector. Following a 5-day activation of natural xenon gas with a 252 Cf (4 ร‚ 10 5 n=s) source, the activities of two gamma ray lines at 164 and 236 keV, from 131m Xe and 129m Xe metastable states, were measured at about 95 and 130 Bq/kg, respectively. We also observed three additional lines at 35, 100 and 275 keV, which decay away within a few days. No long-lifetime activity was observed after the neutron activation.