𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Precise relaxation time measurements of normal-appearing white matter in inflammatory central nervous system disease

✍ Scribed by D. H. Miller; G. Johnson; P. S. Tofts; D. Macmanus; W. I. McDonald


Publisher
John Wiley and Sons
Year
1989
Tongue
English
Weight
370 KB
Volume
11
Category
Article
ISSN
0740-3194

No coin nor oath required. For personal study only.

✦ Synopsis


Precise relaxation time (RT) measurements have been made, with a standard deviation of 3% for TI and T2 in white matter in normal volunteers. This sets an upper limit to the instrumental random errors (imprecision ). Achieving this precision requires careful adjustment and use of the imager. The wide variation in RTs seen by other workers may be in part due to larger instrumental errors. We have measured RTs (both TI and T 2 ) in normal-appearing white matter in 16 normal controls and patients with multiple sclerosis (MS, 18 ), systemic lupus erythematosis (SLE, 16 ) and cerebral sarcoidosis ( 8). Both RTs were significantly higher in MS than in other patient groups and controls ( P < .05), possibly caused by microscopic lesions. T2 was elevated in SLE patients relative to controls and sarcoidosis patients (P < .05), possibly because of microhemorrhages. Lesion RTs were abnormal but more variable and no significant differences between diseases were found.