Polyethylene wear characteristics in vivo and in a knee stimulator
✍ Scribed by Beaulé, P. E. ;Campbell, P. A. ;Walker, P. S. ;Schmalzried, T. P. ;Dorey, F. J. ;Blunn, G. W. ;Bell, C. J. ;Yahia, L'H. ;Amstutz, H. C.
- Book ID
- 102872503
- Publisher
- John Wiley and Sons
- Year
- 2002
- Tongue
- English
- Weight
- 583 KB
- Volume
- 60
- Category
- Article
- ISSN
- 0021-9304
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Validation of a wear simulator requires that the device produce a similar type and amount of wear and particles of a comparable morphology as occurs clinically. Using techniques previously established to compare polyethylene particles from hip simulators to those from retrieved tissues, particles isolated from six revised posterior stabilized knee replacements were characterized and compared to particles generated from the same knee design worn in a knee simulator. The particles produced in the knee simulator were of comparable size but had less variability in their form factor compared to the particles produced in vivo. Comparable wear features were seen on the articulating surfaces in both groups. These results indicate that this knee joint simulator is able to reproduce a baseline type of wear that is similar to that in vivo and should encourage further use of this device to better understand knee component wear and function. © 2002 Wiley Periodicals, Inc. J Biomed Mater Res 60: 411–419, 2002; DOI 10.1002/jbm.10139
📜 SIMILAR VOLUMES
The role that ultrahigh molecular weight polyethylene (UHMWPE) wear at bearing surfaces in total hip and knee arthroplasties plays in the clinical life of these implants cannot be overemphasized. This subject has therefore been attracting enormous research attention on both fundamental and technolog
## Abstract Reduction of wear with highly crosslinked polyethylene (HXLPE) has been reported in __in vitro__ and __in vivo__ studies of total hip prostheses. However, use of HXLPE in total knee prostheses is still controversial. The aim of this study was to compare __in vivo__ polyethylene wear par
Periprosthetic tissue was obtained at revision surgery from 10 posterior cruciate retaining total knee replacement cases (five different manufacturers). The tissues were hydrolyzed and polyethylene particles were isolated from each case. Individual particles were analyzed by scanning electron micros