Piezoelectric crystal biosensor modified with protein A for determination of immunoglobulins
โ Scribed by Muramatsu, Hiroshi.; Dicks, Jonathan M.; Tamiya, Eiichi.; Karube, Isao.
- Book ID
- 120034114
- Publisher
- American Chemical Society
- Year
- 1987
- Tongue
- English
- Weight
- 483 KB
- Volume
- 59
- Category
- Article
- ISSN
- 0003-2700
No coin nor oath required. For personal study only.
โฆ Synopsis
Determlnatlon of lmmunoglobullns was performed by uslng a piezoelectric crystal bhensor system. AT-cut crystals, wlth a baslc resonant frequency of 0 YHr, were modlfled by Immobillring Protein A onto the surface of crystal electrodes wlth ( y-am1nopropyl)trlethoxysllane. The crystal was posi- tloned lnslde a thennostated How cell and measurement of the resonant frequency was performed In constantly flowlng delonlzed/dlstilled water. The resonant frequency shlft resultlng from the afflnlty reactlon of Protein A and human IgG correlated wlth a concentration of human IgG In the range 10a-10-2 mg-mL-'. The analysis pattern of mouse IgG subclasses, obtalned by uslng a stepped gradient of buffer solution, ldentlfied three dlstlnctlve peaks of IgG,, IgG,,, and IgG2b.