Piezoelectric actuator design for MR elastography: implementation and vibration issues
✍ Scribed by Zion Tsz Ho Tse; Yum Ji Chan; Henning Janssen; Abbi Hamed; Ian Young; Michael Lamperth
- Publisher
- Wiley (Robotic Publications)
- Year
- 2011
- Tongue
- English
- Weight
- 329 KB
- Volume
- 7
- Category
- Article
- ISSN
- 1478-5951
- DOI
- 10.1002/rcs.405
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Background
MR elastography (MRE) is an emerging technique for tumor diagnosis. MRE actuation devices require precise mechanical design and radiofrequency engineering to achieve the required mechanical vibration performance and MR compatibility.
Method
A method of designing a general‐purpose, compact and inexpensive MRE actuator is presented. It comprises piezoelectric bimorphs arranged in a resonant structure designed to operate at its resonant frequency for maximum vibration amplitude. An analytical model was established to understand the device vibration characteristics.
Results
The model‐predicted performance was validated in experiments, showing its accuracy in predicting the actuator resonant frequency with an error < 4%. The device MRI compatibility was shown to cause minimal interference to a 1.5 tesla MRI scanner, with maximum signal‐to‐noise ratio reduction of 7.8% and generated artefact of 7.9 mm in MR images.
Conclusions
A piezoelectric MRE actuator is proposed, and its implementation, vibration issues and future work are discussed. Copyright © 2011 John Wiley & Sons, Ltd.