## Abstract The fast acquisition interleaved spin‐echo (FAISE) method is a partial RF echo‐planar technique which utilizes a specific phase‐encode reordering algorithm to manipulate image contrast (Melki __et al., J. Magn. Reson. imaging__ 1:319, 1991). The technique can generate “spin‐echo” like i
Partial RF echo-planar imaging with the FAISE method. II. Contrast equivalence with spin-echo sequences
✍ Scribed by Philippe S. Melki; Ferenc A. Jolesz; Robert V. Mulkern
- Publisher
- John Wiley and Sons
- Year
- 1992
- Tongue
- English
- Weight
- 688 KB
- Volume
- 26
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
The fast acquisition interleaved spin‐echo (FAISE) sequence and its dual‐echo version (DEFAISE) are partial RF echo‐planar methods which utilize a specific phase‐encode reordering algorithm to manipulate T~2~ contrast via an operator‐controlled pseudo‐echo time, pTE. The repetition time, TR, between successive applications of the Carr‐Purcell‐ Meiboom‐Gill (CPMG) echo trains used in FAISE may be reduced to introduce T~1~ weighting. To quantitatively determine the extent to which FAISE T~1~ and T~2~ contrast characteristics agree with spin‐echo methods, signal intensities from FAISE acquisitions were compared with signal intensities from equivalent CPMG acquisitions. In phantoms and in human heads, the contrast Characteristics of FAISE are found to be highly correlated with that obtained with equivalent CPMG sequences. However, conventional SE sequences generally utilize longer echo spacings than employed with FAISE/CPMG. Thus, echo spacing‐dependent mechanisms such as spin‐spin coupling and magnetic susceptibility lead to some differences in contrast between conventional SE and FAISE. Finally, FAISE appears to be more sensitive to magnetization transfer effects than conventional SE sequences since more off‐resonance irradiation is applied to individual slices during multislice acquisitions.
📜 SIMILAR VOLUMES
## Abstract ## Purpose To optimize and assess the feasibility of a single‐shot black‐blood T2‐weighted spin‐echo echo‐planar imaging (SSBB‐EPI) sequence for MRI of the liver using sensitivity encoding (SENSE), and compare the results with those obtained with a T2‐weighted turbo spin‐echo (TSE) seq