✦ Synopsis


Abstract

T~1~ is often ill‐determined. This means that an Ernst angle excitation often cannot be precisely defined for the simple pulse and acquire experiment. Here, published ^31^P T~1~ values of metabolites in human muscle, liver, heart, and brain are archived, some new data on heart and brain added, and overall confidence intervals determined. Strategies for setting the flip‐angle based on the confidence intervals are examined, and an optimum flip‐angle derived which minimizes the signal loss relative to what could have been realized if T~1~ were precisely known. With such optimized pulses, signal loss can be limited to ≤14% for up to a 10‐fold variation in T~1~, with TR ≤ T~1~ The effect that an uncertainty in T~1~ by a factor of two has on the saturation corrected signal is limited to ≤20% in the optimum flip‐angle experiment. Adiabatic B~1~‐independent rotation phase‐cycled (BIRP) excitation pulses are ideal as optimum flip‐angle pulses as they can be prescribed without calibration.


📜 SIMILAR VOLUMES


Concatenated and parallel optimization f
✍ Defeng Wang; Lin Shi; Yi-Xiang J. Wang; Jing Yuan; David K.W. Yeung; Ann D. King 📂 Article 📅 2010 🏛 John Wiley and Sons 🌐 English ⚖ 395 KB

## Abstract Most traditional methods for __T__~1~ map estimation in MRI with fast low‐angle‐shot sequences are aimed at high efficiency by compromising the fitting accuracy. In this paper, the __fundamental__ problem of parameter estimation in fast low‐angle‐shot MRI was re‐examined, and an accurat

Improved T1-weighted two-dimensional MP-
✍ Frederick H. Epstein; John P. Mugler III; James R. Brookeman 📂 Article 📅 1994 🏛 John Wiley and Sons 🌐 English ⚖ 863 KB

## Abstract The recently introduced method of shaping the transient signal evolution in magnetization‐prepared gradient‐echo (MP‐GRE) imaging with variable flip angles has been applied to two‐dimensional (2D) MP‐GRE imaging of the abdomen. The technique was analyzed by using theoretical models and