Optimizing free electron laser incisions at 6.45 μm with computer-assisted surgical techniques and the utilization of a heat-conducting template
✍ Scribed by Jason B. Robbins; Lou Reinisch; Darrel L. Ellis
- Publisher
- John Wiley and Sons
- Year
- 2001
- Tongue
- English
- Weight
- 572 KB
- Volume
- 28
- Category
- Article
- ISSN
- 0196-8092
- DOI
- 10.1002/lsm.1034
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Background and Objective
We sought to optimize laser incisions by evaluating the effects of varying the slit width of a heat‐conducting template and the step size with the Computer‐Assisted Surgical Techniques (CAST) system on free electron laser (FEL) incisions at 6.45 μm.
Study Design/Materials and Methods
Stainless steel calipers were used as a heat‐conducting template on human skin in vitro. The CAST system made the incisions as a series of spot ablations with set step sizes of 50 μm or 1,250 μm. At each step size, incisions were made with or without the calipers, by using varying slit widths. Histologic specimens were analyzed for lateral thermal damage over the entire depth of the incision and over the superficial 150 μm of dermis.
Results
Lateral thermal damage over the superficial 150 μm of dermis was most significantly reduced with the calipers at a slit width approximating the FEL's beam diameter (636 μm ± 100 μm). Also, incisions made with the larger step size (1,250 μm) had significantly less lateral thermal damage over the entire depth of the incision.
Conclusions
The use of a heat‐conducting template with an aperture approximating the FEL's beam diameter and larger step size improved FEL incisions at 6.45 μm. Lasers Surg. Med. 28:162–167, 2001. © 2001 Wiley‐Liss, Inc.