𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Optimized enzyme-linked immunosorbent assay for detection of human and bovine rotavirus in stools: Comparison with electron-microscopy, immunoelectro-osmophoresis, and fluorescent antibody techniques

✍ Scribed by P. C. Grauballe; B. F. Vestergaard; A. Meyling; J. Genner


Publisher
John Wiley and Sons
Year
1981
Tongue
English
Weight
741 KB
Volume
7
Category
Article
ISSN
0146-6615

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Detection of human and bovine rotavirus in stools is described using a double‐antibody sandwich enzyme‐linked immunosorbent assay (ELISA) with polystyrene microtest plates as solid phase, immunoglobulin fraction of rabbit antiserum to rotavirus (human) as catching antibody, and the same reagent labelled with horseradish peroxidase as conjugate. The ELISA has been optimized with regard to simplicity, rapidity, sensitivity, and specificity. In a comparative study, stool specimens from 81 infants and children and 92 neonatal calves with diarrhoea were tested for rotavirus by ELISA, electron microscopy (EM), immunoelectro‐osmorphoresis (IEOP), and fluorescent antibody technique (FA). The relative sensitivity of the different assays for human and bovine rotavirus was: EM 68%, 76%; IEOP 80%, 76%; FA not determined, 85%; and ELISA 86%, 98%, respectively. Less than 1 ng of purified human rotavirus could be detected in ELISA, whereas 100 ng was the minimal amount detected by IEOP. It is concluded that the developed ELISA is a simple, rapid, reliable, and sensitive method for the diagnosis of human and bovine rotavirus infections.