𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Optimization of Folin-Ciocalteu reagent concentration in an automated Lowry protein assay

✍ Scribed by G.M. Oosta; N.S. Mathewson; G.N. Catravas


Publisher
Elsevier Science
Year
1978
Tongue
English
Weight
197 KB
Volume
89
Category
Article
ISSN
0003-2697

No coin nor oath required. For personal study only.

✦ Synopsis


for protein concentration measurement has been automated to permit assay of samples with concentrations from 1 to 400 @g/ml. Calibration with solutions of bovine serum albumin resulted in a nonlinear (quadratic) curve. The quantity of color developed in the assay was found to be strongly dependent on the concentration of the Folin-Ciocalteu phenol reagent. Color yield peaked sharply at a reagent concentration 40% lower than that used in the Lowry procedure. Optimization of the reagent concentration is necessary to obtain maximum sensitivity from the Lowry assay.