An automated robotic method for sample pretreatment was developed for copper determination in environmental and biological materials at the low microgram per gram level. The robotic station performing the preliminary operations as weighing of sample and acid digestion was coupled with the flowinject
Optimization of Flow Injection On-Line Microcolumn Preconcentration of Ultratrace Elements in Environmental Samples prior to Their Spectrochemical Determination
✍ Scribed by Zs. Horváth; A. Lásztity; K. Zih-Perényi; Á. Lévai
- Book ID
- 102617297
- Publisher
- Elsevier Science
- Year
- 1996
- Tongue
- English
- Weight
- 108 KB
- Volume
- 54
- Category
- Article
- ISSN
- 0026-265X
No coin nor oath required. For personal study only.
✦ Synopsis
The preconcentration of some elements such as Cd, Co, Ni, and V(IV) was modeled in the presence of complexing agents such as citrate and oxalate at high Ca, Mg, and sulfate concentrations on iminodiacetic acid/ethyl cellulose (IDAEC), a chelating cellulose. The effect of the species present in the solution was studied after construction of the species distribution curves using critical, estimated, and measured stability constants. The stability constants of the IDAEC chelates were determined potentiometrically. The constants were calculated or estimated using computer programs. The diagrams calculated in homogeneous media were used for optimization of the flow injection on-line preconcentration for analysis of ultratrace metals in the highly mineralized water "Hunyadi."
📜 SIMILAR VOLUMES