Optimal sterilization method for the zirconia/alumina composites used for total hip replacements
โ Scribed by Kwang Woo Nam; Jeong Joon Yoo; Kyung-Hoi Koo; Kang Sup Yoon; Hee Joong Kim
- Book ID
- 102299195
- Publisher
- John Wiley and Sons
- Year
- 2009
- Tongue
- English
- Weight
- 124 KB
- Volume
- 90B
- Category
- Article
- ISSN
- 1552-4973
No coin nor oath required. For personal study only.
โฆ Synopsis
Abstract
Zirconia/alumina composite ceramics have been recently developed for total hip arthroplasty because of their excellent mechanical properties and tribologic characteristics. All such materials used clinically must be easily sterilized, but no report has been issued concerning methods for sterilizing zirconia/alumina composite hip prostheses. Here, we show that 50 kGy of gamma irradiation effectively sterilizes both the surfaces and interiors of these materials. In addition, it was found that the commonly used ethylene oxide and 25โkGy gamma irradiation sterilization methods inadequately sterilize deep inside the femoral head. Moreover, no changes in the chemical or mechanical properties of the composites were noted after exposure to 50โkGy gamma irradiation. We suggest that 50โkGy gamma irradiation is an optimal sterilization method for zirconia/alumina composite total hip replacements. ยฉ 2009 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2009