## Abstract In flow‐imaging experiments with 2‐D Fourier transform sequences, the time difference between phase encoding and readout leads to a potentially misleading displacement artifact. This artifact arises in regions of rapid flow and high shear, and manifests as an intensity distortiion in ad
On the nature and reduction of plaque-mimicking flow artifacts in black blood MRI of the carotid bifurcation
✍ Scribed by David A. Steinman; Brian K. Rutt
- Publisher
- John Wiley and Sons
- Year
- 1998
- Tongue
- English
- Weight
- 1003 KB
- Volume
- 39
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Cardiac-gated black blood MRI of the carotid artery bifurcation in normal human subjects shows signal within the lumen suggesting wall thickening or atherosclerotic plaque. This signal was believed to be artifactual, arising from complex flow patterns present at the carotid bifurcation. Computer simulation of the hemodynamics and black blood multislice image acquisition in a model of the carotid bifurcation showed that these artifacts arise from spins recovering their signal within the slow, recirculating flow of the carotid bulb. The computed hemodynamics also suggested that these artifacts could be minimized or eliminated entirely by gating the acquisition of slices in the most artifact-prone region of the carotid bulb within a 250-ms window after peak systole. Application of these predictions to studies of normal volunteers showed that, in most cases, these flow artifacts in black blood MRI can be eliminated simply by altering the phase of the cardiac cycle to which the image acquisition is gated. The observation that the size and placement of the saturation slabs had little effect on these artifacts suggested that, in those cases in which recirculation persists throughout the cardiac cycle, either inversion-recovery or presaturation within the bulb itself would be required to suppress them.
📜 SIMILAR VOLUMES
## Abstract In this study, a turbo spin‐echo (TSE) based motion‐sensitized driven‐equilibrium (MSDE) sequence was used as an alternative black‐blood (BB) carotid MRI imaging scheme. The MSDE sequence was first optimized for more efficient residual blood signal suppression in the carotid bulb of hea
This study investigated the correlation between changes in hepatic hemodynamics and esophageal variceal pressuremeasured with a noninvasive, pressure-sensitive endoscopic gaugein 37 portal-hypertensive cirrhotic patients receiving propranolol (0.15 mg/kg, intravenously; n = 21) or placebo (n = 16) u
Contrast-enhanced magnetic resonance imaging (MRI) measurement of cerebral perfusion is a diagnostic procedure increasingly gaining access to clinical practice not only in spontaneously breathing patients but also in mechanically ventilated patients. Effects of increased mean airway pressure on cere
## Abstract ## Purpose: To prospectively evaluate whether dose reduction and the application of a prebolus technique can effectively alleviate signal saturation effects in T1 dynamic contrast enhanced (T1‐DCE) magnetic resonance imaging (MRI) data in breast tumors and lead to increased diagnostic