On-line analysis of NH3, CO2, and SO2 diffusion through microporous Teflon
β Scribed by Irina Tarische; Decebal Ciurchea; Patrizio Madau; Liliana Olenic
- Publisher
- John Wiley and Sons
- Year
- 1999
- Tongue
- English
- Weight
- 250 KB
- Volume
- 11
- Category
- Article
- ISSN
- 0895-7533
No coin nor oath required. For personal study only.
β¦ Synopsis
Dynamic responses of self-made potentiometric NH 3 , CO 2 , and SO 2 sensitive gas sensors are followed in the 10 β«5Χβ¬ -5 β«Χβ¬ 10 β«3Χβ¬ M concentration range. By least squares fitting of potential versus time curves, Dk product values at the diffusion of NH 3 , CO 2 , and SO 2 through microporous Teflon are calculated. The obtained results are fairly reflected by the behavior of the sensor. Measurements of potential versus time in mixed solutions of CO 2 and SO 2 with the CO 2 sensitive sensor indicate the presence of the interferent.