On-Line Analysis of Mercury by Sequential Injection Stripping Analysis (SISA) Using a Chemically Modified Electrode
✍ Scribed by Iva Turyan; Thomas Erichsen; Wolfgang Schuhmann; Daniel Mandler
- Publisher
- John Wiley and Sons
- Year
- 2001
- Tongue
- English
- Weight
- 84 KB
- Volume
- 13
- Category
- Article
- ISSN
- 1040-0397
No coin nor oath required. For personal study only.
✦ Synopsis
A highly ¯exible, automatic sequential-injection stripping analysis (SISA) system has been developed and applied for monitoring the levels of mercury in aqueous solutions using a chemically modi®ed electrode. The preparation of the modi®ed electrode comprises spin coating of an ethanolic solution of poly(4-vinylpyridine) and Krypto®x-222 onto a glassy carbon electrode (GCE) followed by cross-linking the polymer. The analysis is based on the anodic stripping voltammetry of mercury using differential pulse voltammetry. A sequence of 36 operations was needed to complete a full cycle of cleaning, calibration and analysis.