𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Occlusion of a large coronary-cameral fistula using the Amplatzer vascular plug in a 2-year old

✍ Scribed by Duraisamy Balaguru; Annette Joseph; Carey Kimmelstiel


Publisher
John Wiley and Sons
Year
2006
Tongue
English
Weight
161 KB
Volume
67
Category
Article
ISSN
1522-1946

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

A 2‐year‐old girl underwent transcatheter occlusion of a fistula. The fistula originated from the right coronary artery and drained into the right atrium. Three interlocking Gianturco coils were initially placed in the distal portion of the fistula resulting in near‐total occlusion of blood flow. Placing a 12‐mm Amplatzer vascular plug obliterated the proximal, dilated portion of the fistula. Complete occlusion with accurate placement was achieved. Advantages and technical implications of the Amplatzer vascular plug in this clinical setting are discussed. © 2006 Wiley‐Liss., Inc.