A method is described for multislice imaging by means of rf pulses which contain a multifrequency excitation spectrum. These pulses label each slice by an individual signal phase and allow the separation of the signals from all the slices by subsequent add/subtract procedures. The image sensitivity
Numerical analysis of multislice mr excitation and inversion with multifrequency selective rf pulses
✍ Scribed by Hans-Peter Hafner; Stephan Müler; Joachim Seelig
- Publisher
- John Wiley and Sons
- Year
- 1990
- Tongue
- English
- Weight
- 744 KB
- Volume
- 13
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Multifrequency selective excitation and inversion were recently described and tested for multislice imaging and multivolume selective spectroscopy (Magn. Reson. Med. 6, 364 (1988), J. Magn. Reson. 76, 155 (1988)). The technique is based on the assumption that a multifrequency rf pulse, a linear superposition of several selective rf pulses with different frequencies, generates a MR signal which can be separated into the spin responses due to each individual frequency. This assumption is investigated theoretically by analyzing the effect of multifrequency selective rf pulses on the magnetization of a homogeneous phantom as a function of slice separation, pulse shape, and rf amplitude using computer simulations of the Bloch equations. It is found that multifrequency selective excitation with sinc pulses–up to eight slices are investigated–and two‐frequency inversion with hyperbolic secant pulses lead to profiles comparable in quality and selectivity to those of conventional single‐frequency pulses.
📜 SIMILAR VOLUMES
## Abstract A weighted averaging scheme with flip angle adaptation is presented for segmented 2D‐selective RF excitations and applied to single‐line segments of a blipped‐planar trajectory. Segments covering the central __k__‐space, i.e. those with significant signal contributions, are averaged mor
## Abstract ## Purpose To selectively visualize the hepatic vein and inferior vena cava (IVC) using three‐dimensional (3D) true steady‐state free‐precession (SSFP) MR angiography with time‐spatial labeling inversion pulse (T‐SLIP), and to optimize the acquisition protocol. ## Materials and Method
## Abstract ## Purpose To evaluate the influence of food intake on portal flow using unenhanced magnetic resonance imaging (MRI). ## Materials and Methods The study population included 29 healthy subjects. A selective inversion recovery tagging pulse was used on the superior mesenteric vein (SMV