Noise reduction in multi-slice arterial spin tagging imaging

✍ Scribed by K. S. St. Lawrence; J. A. Frank; P. A. Bandettini; F. Q. Ye


Publisher
John Wiley and Sons
Year
2005
Tongue
English
Weight
230 KB
Volume
53
Category
Article
ISSN
0740-3194

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Attenuating the static signal in arterial spin tagging (ASSIST) was initially developed for 3D imaging of cerebral blood flow. To enable the simultaneous collection of cerebral blood flow and BOLD data, a multi‐slice version of ASSIST is proposed. As with the 3D version, this sequence uses multiple inversion pulses during the tagging period to suppress the static signal. To maintain background suppression in all slices, the multi‐slice sequence applies additional inversion pulses between slice acquisitions. The utility of the sequence was demonstrated by simultaneously acquiring ASSIST and BOLD data during a functional task and by collecting resting‐state ASSIST data over a large number of slices. In addition, the temporal stability of the perfusion signal was found to be 60% greater at 3 T compared to 1.5 T, which was attributed to the insensitivity of ASSIST to physiologic noise. Magn Reson Med 53:735–738, 2005. Published 2005 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES


Correction of translational motion artif
✍ L.-H. Zang; J. Fielden; J. Wilbrink; A. Takane; H. Koizumi 📂 Article 📅 1993 🏛 John Wiley and Sons 🌐 English ⚖ 983 KB

## Abstract In this paper, we describe a method for detecting and correcting in‐plane bulk translational motion in multislice spin‐echo imaging using self‐calibration and postprocessing. A constant phase encoding offset between slices is used to evenly spread out the low spatial frequency echoes to

Reduction of errors in ASL cerebral perf
✍ Jack A. Wells; David L. Thomas; Martin D. King; Alan Connelly; Mark F. Lythgoe; 📂 Article 📅 2010 🏛 John Wiley and Sons 🌐 English ⚖ 466 KB

## Abstract In this work, the performance of image de‐noising techniques for reducing errors in arterial spin labeling cerebral blood flow and arterial transit time estimates is investigated. Simulations were used to show that the established arterial spin labeling cerebral blood flow quantificatio