## Abstract A battery powered portable monitor based on flow injection potentiometry has been developed weighing 1.8 kg, constructed in a carry‐case and applicable for remote site monitoring. Portability has been achieved using a light weight rechargeable Ni‐Cd battery‐pack (7.2 V). The flow inject
Nitrate ion-selective electrode as reference electrode for flow-injection analysis
✍ Scribed by Julian Alonso-Chamarro; Jordi Bartroli; Cecília Jimenez
- Publisher
- Elsevier Science
- Year
- 1992
- Tongue
- English
- Weight
- 387 KB
- Volume
- 261
- Category
- Article
- ISSN
- 0003-2670
No coin nor oath required. For personal study only.
✦ Synopsis
The development of a flow-mJectlon system for pH determmation usmg a pH Ion-sensltwe field-effect transistor m conJunctlon Hnth a tubular flow-through nitrate Ion-selectwe electrode (ISE) as a reference electrode 1s descrtbed With this system, the hnear workmg range extends from pH 1 to 13 wth a sensltltlvlty of 56 mV per pH umt and a relative standard dewatron of 0 4% A stable baseline and good response are obtained by placmg the tubular ISE m a closed branch of the mam channel
📜 SIMILAR VOLUMES
A fast-response flow cell Hrlth ion-selectwe electrodes was constructed for contmuous-flow measurements m au-segmented streams without debubblmg The flow system was optmuzed for fluonde and chlonde detennmatlons With a flow-rate of 4 ml mm-' a samphng rate of 480 h-' was achieved for chlonde determm
A flow analyser with enhanced samplmg frequency combmmg the features of flow-nuectlon and contmuous-flow analysis IS described Parameters of the analyser were optnmzed to ensure the highest rate of analysis allowed by the bubble-through flow cell wth fluoride-selectnre electrodes Fluonde was determm